Flame Photometer

Flame Photometer

0 TO 199.9ppm, with Na, K, Li, Ca and Ba filters
CodeCE215
British StandardBS4550-21
Other StandardsEN196-21
Approx. Weight (kg)8
Tariff90248000
Click to save page as PDF Save Page to PDF
addthis
or
Call Impact +44 (0)1294 602626

Single channel emission flame photometer for routine determination of Na, K, Ca, Li and Ba.
  • Digital readout
  • Switchable decimal point
  • Narrow band filters on rotating discs
  • Electronic ignition of flame
  • Electronic flame failure protection
  • Automatic regulation of air supply
  • Simple maintenance and cleaning
  • Recorder output
Specification
Range 0 to 199.9ppm
Limits of detection Na 0.2ppm
K 0.2ppm
Li 0.25ppm
Ca 15ppm
Ba 30ppm
Reproducibility <1% coefficient of variation for 20 consecutive samples using 10ppm Na set to read 50
Linearity < 2% error when 3ppm Na/K and 5ppm Li are set to read 100
Stability < 2% drift over 5min when continuously aspirating 10ppm sample set to 50.0. Specificity Interference from Na/K and Li equal in concentration to test element will be <0.5%
Recorder output Nominal 1.00V for a reading of 100.0
Electrical supply 90-125V or 190-250V @50/60Hz
Air supply Moisture and oil free, 6 litre/min @ 14psi
Fuel Propane, butane, natural gas or LPG
Size (w x d x h) 420 x 360 x 300mm
Weight 8kg
Services required
  1. Electrical
    230V 50Hz single phase
  2. Air
    Moisture and oil free 6 litres/minute at 1Kg/cm2
  3. Gas
    Propane, butane or natural gas (specify correct regulator when ordering)

Supplied with Na, K, Li, Ca and Ba filters, fuel and air connectors, instructions and service manual.
Requires but does not include a gas regulator and an air compressor of the correct type.
230V 50Hz 1ph

Click Click to view a PDF file of the User Guide

A user guide is available to download from the Impact web site www.impact-test.co.uk